BioSpa 8 Automated Incubator
Technical Details
General
Microplate types |
6- to 1536-well standard height microplates, with or without lids. Plate height range: 7.6 mm to 25.4 mm |
Other labware supported | Cell culture dishes (60 mm and 35 mm) |
Lidded plate handling | Robotic arm moves plate to and from BioSpa 8 and connected instrument; handles de-lidding and re-lidding plates. |
Microplate capacity | Up to 8 microplates (or other labware) |
Air filter | User-replaceable HEPA filter |
Compatible instruments | Cytation 7, Cytation 5, Cytation 1, Synergy Neo2, Synergy H1, Epoch 2, EL406, 405 TS, 405 LS, MultiFlo FX |
Interfacing capacity | 1 or 2 devices: Reader/imager only, washer/dispenser only, or both |
Software |
BioSpa Automated Incubator Software:
Provides programming interface for BioTek detection and liquid handling devices Allows user notification (text or email) of events and/or errors in the system Provides control, monitoring and logging of: Allows long-term uninterrupted runs |
CO2 and O2 control (option) |
Range: 0 - 20% (CO2); 1 - 19% (O2) Control Resolution: +0.1% (CO2 and O2) Stability: +0.2% at 5% CO2; +0.2% at 1% O2 |
Temperature Control |
Range: to 45 °C Control resolution: 0.1 °C Uniformity: + 0.5 °C at 37 °C |
Humidity (option) |
rH: 80 to 95% (lidded plates and 5% CO2) Source: Removable water pan Water level sensor: Low water level alert |
Recommended Computer System Requirements |
|
Physical Characteristics
Weight | < 85 lbs (39 kg) (without shipping components and an empty water pan) |
Dimensions |
With standard base pan extension: 29.5" L x 22" W x 19"H (75 cm x 55 cm x 48 cm) With fluidics base pan extension: 39" L x 22" W x 19" H (99 cm x 55 cm x 48 cm) |
Regulatory
Regulatory | CE and TUV marked, RoHS compliant. |